A low-temperature chemical sterilization method using ethylene oxide gas. It is effective for sterilizing heat-sensitive and moisture-sensitive materials like plastics, electronic devices, and optical instruments. The process involves alkylation, where EtO disrupts the DNA and proteins of microorganisms, preventing their replication. It requires careful control of gas concentration, temperature, humidity, and exposure time.

Ethylene oxide (C2H4O) is a colorless, flammable, and highly reactive gas. Its effectiveness as a sterilant lies in its ability to act as a potent alkylating agent. The process works at a molecular level. The strained three-membered ring of the ethylene oxide molecule readily opens and reacts with nucleophilic groups in essential cellular components. It primarily targets sulfhydryl (-SH), amino (-NH2), carboxyl (-COOH), and hydroxyl (-OH) groups found in proteins and nucleic acids (DNA and RNA). By adding an alkyl group to these sites, EtO permanently modifies the molecules, blocking reactive sites, causing cross-linking, and ultimately preventing microbial metabolism and replication. This mechanism is effective against all forms of microorganisms, including highly resistant bacterial spores. A typical EtO sterilization cycle is complex and must be carefully controlled. It consists of several phases: preconditioning (adjusting temperature and humidity), gas injection (introducing the EtO), exposure or dwell time (holding the load under specified conditions for several hours), and aeration (removing residual EtO gas). Key parameters are gas concentration (typically 400–1200 mg/L), temperature (usually 37–63 °C), relative humidity (40–80%), and exposure time (1–6 hours). The lower temperature range makes it ideal for materials that would be damaged by the heat of autoclaving, such as polymers, electronics, and medical devices with complex designs. However, EtO is toxic, carcinogenic, and explosive, requiring stringent safety protocols and a lengthy aeration period to ensure residual gas is removed from the sterilized items to safe levels before they can be used.

While ethylene oxide was first synthesized in 1859 by Charles-Adolphe Wurtz, its potent biological activity was not fully realized until much later. It was patented as a fumigant for killing insects by Lloyd L. Schrader in the 1920s. Its application as a sterilizing agent for medical supplies was pioneered in the 1940s and became widespread in the 1950s, particularly with the rise of disposable plastic medical devices. The U.S. military reportedly developed the process for sterilizing equipment. This innovation was revolutionary because it filled a critical gap; before EtO, there was no reliable method for sterilizing the growing number of heat-sensitive instruments and supplies that were becoming essential to modern medicine. It enabled the mass production and use of pre-packaged, sterile, single-use devices, which have significantly reduced the risk of healthcare-associated infections.

UNESCO Nomenclature: 2401
– Microbiology

Type

Chemical Process

Disruption

Revolutionary

Utilisation

Widespread Use

Precursors

  • discovery of ethylene oxide
  • understanding of chemical disinfection and alkylating agents
  • need for sterilizing materials incompatible with heat or steam

Applications

  • sterilization of single-use medical devices (e.g., syringes, catheters)
  • processing of spices to reduce microbial load
  • sterilization of delicate surgical instruments
  • archival and museum artifact preservation

Brevets :

QUE

Potential Innovations Ideas

!niveaux !!! Adhésion obligatoire

Vous devez être membre de l'association pour accéder à ce contenu.

S’inscrire maintenant

Vous êtes déjà membre ? Connectez-vous ici
Related to: ethylene oxide, eto, chemical sterilization, low-temperature, alkylation, medical devices, plastics, toxicity, aeration, heat-sensitive.

Laisser un commentaire

Votre adresse e-mail ne sera pas publiée. Les champs obligatoires sont indiqués avec *

DISPONIBLE POUR DE NOUVEAUX DÉFIS
Mechanical Engineer, Project, Process Engineering or R&D Manager
Développement de produits efficace

Disponible pour un nouveau défi dans un court délai.
Contactez-moi sur LinkedIn
Plastic metal electronics integration, Design-to-cost, GMP, Ergonomics, Medium to high-volume devices & consumables, Lean Manufacturing, Regulated industries, CE & FDA, CAD, Solidworks, Lean Sigma Black Belt, medical ISO 13485

Nous recherchons un nouveau sponsor

 

Votre entreprise ou institution est dans le domaine de la technique, de la science ou de la recherche ?
> envoyez-nous un message <

Recevez tous les nouveaux articles
Gratuit, pas de spam, email non distribué ni revendu

ou vous pouvez obtenir votre adhésion complète - gratuitement - pour accéder à tout le contenu restreint >ici<

Historical Context

(if date is unknown or not relevant, e.g. "fluid mechanics", a rounded estimation of its notable emergence is provided)

Related Invention, Innovation & Technical Principles

Retour en haut

Vous aimerez peut-être aussi